Accuracy of Gram-stained Smears as Screening Tests for Neisseria gonorrhoeae: A Systematic Review and Meta-analysis

Authors

  • Mohamed Abdalla Ali Khalid Laboratory Department, Kamal Medical Polyclinic. Riyadh 11912, Saudi Arabia

DOI:

https://doi.org/10.54844/cai.2021.0014

Abstract

Background and Objective: A total of 86.9 million persons worldwide are infected with Neisseria gonorrhoeae (Ng). Although Gram-stained smears (GSS) provide a timeand cost-saving alternative to conventional laboratory tests, their global uptake partly depends on their performance. This study aimed to meta-analyze the diagnostic accuracy of GSS to screen for Ng. Materials and Methods: A literature search was conducted using the MEDLINE (1980 to 2020). Studies were included if they employed GSS to detect Ng in humans and compared the results with reference tests. Results: Eleven studies were reviewed and meta-analyzed and stratified by specimen type (Gram-stained urethral smears and Gram-stained endocervical, urethral swabs and urine smears.) and reference test type (culture method or NAAT). Sensitivity was similarly high in GSS versus NAAT (93% [CI, 64% to 99 %]) and GSS versus culture methods (87% [CI, 74% to 94%]), followed by Gram-stained urethral smears (97% [95% CI, 86% to 100%]) and Gram-stained endocervical, urethral swabs and urine smears (81% [CI, 67% to 90%]). Specificity was also high in GSS versus culture methods (98% [CI, 95% to 100%]) and GSS versus NAAT (94% [CI, 73% to 99%]), followed Gram-stained endocervical, urethral swabs and urine smears (98% [CI, 93% to 99%]) and Gram-stained urethral smears (96% [CI, 78% to 99%]). Conclusions: Data suggest that GSS have the highest accuracy when investigated against reference culture methods, and Gram-stained urethral smears have the highest accuracy, followed by Gram-stained endocervical, urethral swabs and urine smears. Given their accuracy, convenience, and quick turnaround time, GSS may be useful in expanding first-line screening Ng.

References

Młynarczyk-Bonikowska B, Majewska A, Malejczyk M, Młynarczyk G, Majewski S. Multiresistant Neisseria gonorrhoeae: a new threat in second decade of the XXI century. Med Microbiol Immunol 2020;209:95–108 DOI: https://doi.org/10.1007/s00430-019-00651-4

Rubin DHF, Ross JDC, Grad YH. The frontiers of addressing antibiotic resistance in Neisseria gonorrhoeae. Translational Research : the Journal of Laboratory and Clinical Medicine 2020;220:122–37 DOI: https://doi.org/10.1016/j.trsl.2020.02.002

Unemo M, Seifert HS, Hook EW, Hawkes S, Ndowa F, Dillon J-AR. Gonorrhoea. Nat Rev Dis Prim [Internet] 2019;5:79.

Meyer T, Buder S. The Laboratory Diagnosis of Neisseria gonorrhoeae: Current Testing and Future Demands. Pathogens 2020;9:91. DOI: https://doi.org/10.3390/pathogens9020091

Qian S, Foster R, Bourne C, Vickers T, McIver R, McNulty A. Neisseria gonorrhoeae positivity in clients presenting as asymptomatic contacts of gonorrhoea at a sexual health centre. Sex Health 2020;17:187–191. DOI: https://doi.org/10.1071/SH19091

Thairu Y, Nasir I, Usman Y. Laboratory perspective of gram staining and its significance in investigations of infectious diseases. Sub-Saharan African J Med 2014;1:168–174. DOI: https://doi.org/10.4103/2384-5147.144725

Papp JR, Schachter J, Gaydos CA , Van Der Pol B. Recommendations for the laboratory-based detection of Chlamydia trachomatis and Neisseria gonorrhoeae—2014. MMWR Recomm reports Morb Mortal Wkly report Recomm reports/Centers Dis Control. 2014;63:1.

Whiting PF, Rutjes AWS, Westwood ME, Mallett S, Deeks JJ, Reitsma JB, et al. QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies. Ann Intern Med 2011;155:529–536. DOI: https://doi.org/10.7326/0003-4819-155-8-201110180-00009

Harbord RM, Whiting P. Meta-analysis of diagnostic accuracy. In: Sterne JA, ed. Meta-analysis in Stata: An Updated Collection From the Stata Journal. College Station, TX: Stata Pr; 2009.

Jaeschke R, Guyatt GH, Sackett DL. Users’ guides to the medical literature. III. How to use an article about a diagnostic test. B. What are the results and will they help me in caring for my patients? The Evidence-Based Medicine Working Group. JAMA 1994;271:703–707. DOI: https://doi.org/10.1001/jama.271.9.703

Bhargava A, Bala M, Singh V, Joshi NC, Kakran M, Puri P, et al. How reliable is microscopy and culture for the diagnosis of gonorrhea? An 11-year experience from India. Sex Transm Dis 2017;44:111–113. DOI: https://doi.org/10.1097/OLQ.0000000000000550

Situ SF, Ding CH, Nawi S, Johar A, Ramli R. Conventional versus molecular detection of Chlamydia trachomatis and Neisseria gonorrhoeae among males in a sexually transmitted infections clinic. Malays J Pathol 2017;39:25–31.

Hananta IPY, Van Dam AP, Bruisten SM, Van Der Loeff MFS, Soebono H, Christiaan De Vries HJ. Value of light microscopy to diagnose urogenital gonorrhoea: A diagnostic test study in Indonesian clinic-based and outreach sexually transmitted infections services. BMJ Open 2017;7:1–12. DOI: https://doi.org/10.1136/bmjopen-2017-016202

Goodhart ME, Ogden J, Zaidi AA, Kraus SJ. Factors affecting the performance of smear and culture tests for the detection of Neisseria gonorrhoeae. Sex Transm Dis 1982;9:63–69. DOI: https://doi.org/10.1097/00007435-198204000-00002

Juchau SV., Nackman R, Ruppart D. Comparison of gram stain with DNA probe for detection of Neisseria gonorrhoeae in urethras of symptomatic males. J Clin Microbiol 1995;33:3068–3069. DOI: https://doi.org/10.1128/jcm.33.11.3068-3069.1995

Taylor SN, Dicarlo RP, Martin DH. Comparison of methylene blue/gentian violet stain to gram’s stain for the rapid diagnosis of gonococcal urethritis in men. Sex Transm Dis 2011;38:995–996. DOI: https://doi.org/10.1097/OLQ.0b013e318225f7c2

Goh BT, Varia KB, Ayliffe PF, Lim FK. Diagnosis of gonorrhea by gram-stained smears and cultures in men and women: role of the urethral smear. Sex Transm Dis 1985;12:135–139. DOI: https://doi.org/10.1097/00007435-198507000-00008

D’Angelo LJ, Mohla C, Sneed J, Woodward K. Diagnosing gonorrhea. A comparison of standard and rapid techniques. J Adolesc Heal Care 1987;8:344–348. DOI: https://doi.org/10.1016/0197-0070(87)90006-4

Orellana Miguel MA, Gómez-Lus ML, Lora D. Sensitivity of Gram stain in the diagnosis of urethritis in men. Sex Transm Infect 2012;88:284–287. DOI: https://doi.org/10.1136/sextrans-2011-050150

Bartelsman M, Straetemans M, Vaughan K, Alba S, Van Rooijen MS, Faber WR, et al. Comparison of two Gram stain point-of-care systems for urogenital gonorrhoea among high-risk patients: Diagnostic accuracy and cost-effectiveness before and after changing the screening algorithm at an STI clinic in Amsterdam. Sex Transm Infect 2014;90:358–362. DOI: https://doi.org/10.1136/sextrans-2013-051500

Borg S, Daniel J, Forsyth S. Is urethral microscopy useful in symptomatic male patients without urethral discharge? Int J STD AIDS 2019;30:79–81. DOI: https://doi.org/10.1177/0956462418793511

Ng LK, Martin IE. The laboratory diagnosis of Neisseria gonorrhoeae. Can J Infect Dis 2005;16:15–25. DOI: https://doi.org/10.1155/2005/323082

Whiting P, Rutjes AWS, Reitsma JB, Glas AS, Bossuyt PMM, Kleijnen J. Sources of variation and bias in studies of diagnostic accuracy: a systematic review. Ann Intern Med 2004;140:189–202. DOI: https://doi.org/10.7326/0003-4819-140-3-200402030-00010

Verma R, Sood S. Gonorrhoea diagnostics: An update. Indian J Med Microbiol 2016;34:139–145. DOI: https://doi.org/10.4103/0255-0857.180278

Budkaew J, Chumworathayi B, Pientong C, Ekalaksananan T. Conventional culture versus nucleic acid amplification tests for screening of urethral Neisseria gonorrhea infection among asymptomatic men who have sex with men. Pragmatic Obs Res 2017;8:167–73. DOI: https://doi.org/10.2147/POR.S137377

Ghanem KG , Hynes NA , Deputy D, Bloom A. Clinical manifestations and diagnosis of Neisseria gonorrhoeae infection in adults and adolescents. UpToDate; 2019.

Unemo M, Seifert HS, Hook EW, Hawkes S, Ndowa F, Dillon J-AR. Gonorrhoea. Nat Rev Dis Prim 2019;5:79. DOI: https://doi.org/10.1038/s41572-019-0128-6

Whittles LK, White PJ, Didelot X. A dynamic power-law sexual network model of gonorrhoea outbreaks. PLoS Comput Biol 2019;15:e1006748–e1006748. DOI: https://doi.org/10.1371/journal.pcbi.1006748

Jordan SJ, Schwebke JR, Aaron KJ, Van Der Pol B, Hook 3rd EW. Meatal Swabs Contain Less Cellular Material and Are Associated with a Decrease in Gram Stain Smear Quality Compared to Urethral Swabs in Men. J Clin Microbiol 2017;55:2249–2254. DOI: https://doi.org/10.1128/JCM.00423-17

Raffe S, Soni S. Diagnostic tests for sexually transmitted infections. Medicine Baltimore 2018;46:277–282. DOI: https://doi.org/10.1016/j.mpmed.2018.02.008

Sarier M, Sepin N, Duman I, Demir M, Hizel A, Göktaş Ş, et al. Microscopy of Gram-stained urethral smear in the diagnosis of urethritis: Which threshold value should be selected? Andrologia 2018;50:1–4. DOI: https://doi.org/10.1111/and.13143

Published

2022-03-25

How to Cite

1.
Khalid MAA. Accuracy of Gram-stained Smears as Screening Tests for Neisseria gonorrhoeae: A Systematic Review and Meta-analysis. Community Acquir Infect. 2022;9. doi:10.54844/cai.2021.0014

Issue

Section

Original Articles

Downloads

Download data is not yet available.
Original Article

Accuracy of gram-stained smears as screening tests for Neisseria gonorrhoeae: a systematic review and meta-analysis


Mohamed Abdalla Ali Khalid

Laboratory Department, Kamal Medical Polyclinic. Riyadh 11912, Saudi Arabia


Background and Objective: A total of 86.9 million persons worldwide are infected with Neisseria gonorrhoeae (Ng). Although Gram-stained smears (GSS) provide a timeand cost-saving alternative to conventional laboratory tests, their global uptake partly depends on their performance. This study aimed to meta-analyze the diagnostic accuracy of GSS to screen for Ng. Materials and Methods: A literature search was conducted using the MEDLINE (1980 to 2020). Studies were included if they employed GSS to detect Ng in humans and compared the results with reference tests. Results: Eleven studies were reviewed and meta-analyzed and stratified by specimen type (Gram-stained urethral smears and Gram-stained endocervical, urethral swabs and urine smears.) and reference test type (culture method or NAAT). Sensitivity was similarly high in GSS versus NAAT (93% [CI, 64% to 99 %]) and GSS versus culture methods (87% [CI, 74% to 94%]), followed by Gram-stained urethral smears (97% [95% CI, 86% to 100%]) and Gram-stained endocervical, urethral swabs and urine smears (81% [CI, 67% to 90%]). Specificity was also high in GSS versus culture methods (98% [CI, 95% to 100%]) and GSS versus NAAT (94% [CI, 73% to 99%]), followed Gram-stained endocervical, urethral swabs and urine smears (98% [CI, 93% to 99%]) and Gram-stained urethral smears (96% [CI, 78% to 99%]). Conclusions: Data suggest that GSS have the highest accuracy when investigated against reference culture methods, and Gram-stained urethral smears have the highest accuracy, followed by Gram-stained endocervical, urethral swabs and urine smears. Given their accuracy, convenience, and quick turnaround time, GSS may be useful in expanding first-line screening Ng.


Address for correspondence:

Mohamed Abdalla Ali Khalid, Laboratory Department, Kamal Medical Polyclinic. Riyadh, Saudi Arabia. E-mail: Maak16@hotmail.com

Submitted: 15-04-2021 Revised: 09-06-2021

Accepted: 21-07-2021 Published: 25-03-2022

For reprints contact: reprints@sppub.org



Neisseria gonorrhoeae (Ng) is an etiologic agent of gonorrhea, one of the most common sexually transmitted diseases caused by bacteria,[1] with an estimated global annual incidence of 86.9 millionin adults. Gonorrhea can present as urethritis in men, cervicitis, or urethritis in women, and in extragenital sites (pharynx, rectum, conjunctiva, and, rarely, systemically) in both sexes.[2,3] The ideal laboratory test for the detection of Ng should be sensitive, specific, easy to use, rapid, and affordable.[4] In the United States, the Center for Disease Control and Prevention (CDC) recommends using culture andnucleic acid amplification tests (NAAT) for the diagnosis and detection of Ng in genital, rectal, or pharyngeal secretions.[3,4] These tests have high sensitivity and specificity, but the results are not available until days after testing. This potentially leads to ongoing transmission due to treatment delay or loss to follow-up.[5] A presumptive gonorrhea diagnosis can also be made based on light microscopic detection of the bacterium in Gram-stained smears (GSS). This enables immediate treatment, thus preventing ongoing transmission and/or loss to follow-up.[6] Given the high absolute burden of high prevalence of Ng globally and the high costs of NAAT and culture methods, GSS testing can reduce the burden of cost considerably, thereby reducinggonorrhea prevalence. To address this knowledge gap, this study reviewed evidence on the diagnostic performance of GSS inscreening for gonorrhea.


This review focuses on the diagnostic accuracy variables (sensitivity, specificity, likelihood ratios [LRs], and diagnostic odds ratios [DORs]) of GSSthat are used to screen for Ng in urethral swab andendocervical swab specimens. The author evaluated the studies conducted worldwide in adults regardless of their risk profile in all study settings (laboratoryor field-based) and all study designs (cross-sectional studies and case control). The Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines for reporting synthesiswere followed.


Data Sources and Searches

The searchwas conducted using MEDLINE (via PubMed) from 1980 to 2020. The last search was conducted on June 1, 2020. An example is MEDLINE search string (restricted to humans only): (“Cervical smear for gonococcal” [MeSH] OR “urethral smear for gonococcal” [MeSH] OR “diagnosis of gonorrhea” [MeSH]) AND (“Gram-stained smears for gonorrhea”).


Study Selection

This review includes both abstracts and full-text articles of studies conducted using adult humans, with sufficient raw data to recreate 2 × 2 diagnostic tables. Articles were not excluded on the basis of study location or study design. However, non-English articles, studies on the prevalence or accuracy of laboratory-based tests, and those missing relevant information on the index test (GSS) were excluded. Manufacturers’ reports were not included in this review because they usually provide inadequate details on study conduct, have overt conflicts of interest, often provide accuracy estimations without CIs, and exclude important methodological details in study design, patient populations, and samples. Figure 1 is a flowchart of the search. The author conducted the searches and screened the articles for eligibility. After the initial identification of all studies and deletion of duplicates, a preliminary screening of 937 articles was performed, based on the titles and abstracts. Of these, 26 were considered for full-text review, out of which 10 were included in the study.


Data Extraction and Quality Assessment

The author abstracted data using a prepiloted form and critiqued the quality of the studies. Data were extracted on the characteristics of the study population, including sampling strategies (purposive or consecutive random sampling), sample size, inclusion and exclusion criteria, specimen tested (endocervical and/or urethral swab), whether the test was a gram stain the reference standard, funding sources, and any reported conflicts of interest. The authoralso extracted raw data of true positive, true negative, false positive, and false negative results and items necessary for the assessment of the study quality.


In this study, the GSS method was considered, per CDC recommendations[7]: First, swabs were rolled onto clean glass slides and smeared over an area of less than 1 cm2. Second, the smears were heat-fixed and gram-stained. Finally, the stained smears were examined using a light microscope under oil immersion (1000× magnification) for gram-negative diplococci and their spatial relationship to polymorphonuclear leukocytes.


Culture techniques followed by confirmation of isolates by biochemical, enzymatic, serologic, or nucleic acid testing, for example, carbohydrate utilization, rapid enzyme substrate tests, serologic methods such as coagglutination, or fluorescent antibody tests, are considered reference standards on the basis of CDC recommendations.[7]


In this study, the methodological and reporting qualities of studies were assessed using the QUADAS-2 (Quality Assessment of Diagnostic Accuracy Studies 2) tool.[8] The QUADAS-2 checklist assesses potential bias in studies with respect to patient selection, index test, reference test, and patient flow.[8] In assessing the quality of thestudies, we also focused on the reference standards used and any reported conflict of interest.


Data Synthesis and Analysis

All statistical analyseswere carried out using Intercooled Stata, version 15 (StataCorp, College Station, Texas, USA). For the meta-analysis of the estimates of accuracy, the researcher usedthe bivariate model, which assumes that the measures ofsensitivity and specificity from a study are negatively correlatedand that the logit transformations of sensitivity andspecificity have a bivariate normal distribution.[9] The sensitivity, specificity, positive LR, negative LR, and DOR were calculated. The LRs of a testinform the pretest probability ofthe disease and provide a post-test probability. A positive LR higher than 5 and negative LR less than 0.2 providestrong diagnostic evidence.[10]


Before the meta-analysis, we stratified the studies into four subgroupsbased on the specimen tested and whether the reference testwas a culture method or NAAT. Because the data were insufficient for all tests and all of the tests under investigation were GSS, we stratified evidence into two subgroups onthe same basis: Gram-stained urethral smearsand Gram-stained endocervical smears.


Characteristics of Studies

Author search returned 973 reports, of which 11 satisfied all of the inclusion criteria Figure 1. Table 1[11,12,13,14,15,16,17,18,19,20,21] shows the study characteristics. A total of 11 studies were reviewed and analyzed.


Figure 1
Figure 1: Data extraction and quality assessment


Table 1: Characteristics of reviewed studies
Author, Year (Reference) Location Sample size Study design Reference standard Specimen
Bhargava, 2017[11] India 10,964 Note Culture Endocervical and urethral swab
Goodhart, 1982[14] United States 401 Cross-sectional Culture Endocervical and urethral swab
Taylor, 2011[16] United States 307 Cross-sectional Culture urethral swab
Goh, 1985[17] United States 27,600 Cross-sectional Culture Endocervical and urethral swab
D’ANGELO, 1987[18] United States 419 Cross-sectional Culture Endocervical and urethral swab
Orellana, 2007[19] Spain 491 Cross-sectional Culture urethral swab
Bartelsman, 2011[20] Netherlands 22,707 Cross-sectional Culture Endocervical and urethral swab
Borg, 2017[21] United kingdom 180 Retrospective audit NAAT urethral swab
JUCHAU, 1995[15] United States 7,429 Cross-sectional Gen-Probe urethral swab
Hun, 2017[12] Malaysia 95 Cross-sectional Xpert® CT/NG assay urethral swab, urine
Hananta, 2017[13] Indonesia 632 Post hoc, exploratory analysis Ng-PCR Rotorgene System Endocervical and urethral swab


Of the 11studies, 3 (27 %) were conducted in developing settings[11,12,13] and 8 (73%) were conducted in developed settings.[14,15,16,17,18,19,20,21] Sample sizes ranged from 95 to 27600 persons.


Study Quality

Eight studies (73%)[12,14,15,16,17,18,19,20] were cross-sectional (assessed using the QUADAS-2 checklist). Seven studies (64%)[11,14,16,17,18,19,20] used a CDC-recommended reference standard (culture method), whereas the remaining four studies (36%) used molecular tests (NAAT, Gen-Probe, Ng-PCR Rotorgene system, or Xpert® CT/NG assay) as the reference standard. All the research groups administered the same reference test to all patients, thus avoiding partial or differential verification bias (Figure 2).


Figure 2
Figure 2: Quality assessment of diagnostic accuracy studies assessments


Five studies (45%)[11,12,15,20,21] reported a financial relationship with or received funding, 8 (73%)[11,12,13,14,15,16,17,18] omitted disclosure of conflicts of interest, 3 (27 %)[19,20,21] explicitly declared no conflict of interest, and 1[20] reported receiving tests inkind from manufacturers but no conflict of interest.


Results Pooled by Subgroup

Table 2 reports estimates of sensitivity and specificity from each study. Table 3 pooled estimates for each subgroup.


Table 2: Sensitivity and specificity from each study
Study Sensitivity (95% CI), % Specificity (95% CI), %
Bhargava/2017 95 (93-97) 99 (99-99)
Goodhart/1982 70 (63-76) 85 (79-89)
Tayor/2011 99 (93-100) 99 (97-100)
Goh /1985 90 (88-91) 98 (97-98)
D’ANGELO/1987 56 (42-69) 99 (98-100)
Orellana/2007 80 (61-92) 90 (87-93)
Bartelsman/2011 86 (83-87) 100 (100-100)
Borg/2017 91 (76-98) 64 (55-71)
JUCHAU/1995 100 (99-100) 100 (99-100)
Hun/2017 90 (74-98) 95 (87-99)
Hananta/2017 53 (43-62) 89 (86-92)


Table 3: Results of Meta-analysis, by specimen and reference standard subgroup
Subgroup Pooled Sensitivity (95% CI), % Pooled Specificity (95% CI), % Positive LR (95% CI), % Negative LR (95% CI),% DOR (95% CI),%
GSS verse culture methods 87 (74-94) 98 (95-100) 55.9 (16-196) 0.13 (0.06-0.28) 417 (78-2226)
GSS verse NAAT 93 (64-99) 94 (73-99) 16.2 (2.7-96) 0.07 (0.01-0.52) 225 (6-7842)
Gram-stained urethral smears 97 (86-100) 96 (78-99) 25.9 (3.7-180.7) 0.03 (0.00-0.17) 901 (24-33445)
Gram-stained endocervical, urethral swabs and urine smears 81 (67-90) 98 (93-99) 40.8(11.5-143.8) 0.19 (0.10-0.36) 215 (41-1126)
DOR: diagnostic odds ratio; LR: likelihood ratio; GSS: Gram-stained smears; NAAT: Nucleic acid amplification tests


Diagnostic performance of GSS versusculture methods

The tests investigated in this subgroup were GSS compared withthe reference culture method. Among the seven data points, the pooled sensitivity was 87% (95% CI, 74%-94%) and the pooled specificity was 98% (CI, 95%-100%). The positive LR was (55.9% [CI, 16%-196%]), negative LR (0.13 [CI, 0.06%-0.28%]), and DOR (417 [CI, 78%-2226%]).


Diagnostic performance of GSS versusNAAT

The tests in this subgroup were GSS and NAAT (Gen-Probe, Ng-PCR Rotor gene system, and Xpert® CT/NG assay). Among the fourdatapoints, the pooled sensitivity was 93% (CI, 64%-99 %) and the pooled specificity was 94% (CI, 73%-99%). The positive LR was 16.2 (CI, 2.7%-96%), the negative LR was 0.07 (CI, 0.01%-0.52%), andthe DOR was 225 (CI, 6%-7842%).


Diagnostic performance of Gram-stained urethral smears

Among the four data points, thepooled sensitivity was 97% (CI, 86%-100%) andthe pooled specificity was 96% (CI, 78%-99%). The DOR for this subgroup was 901(CI, 24%-33445%). The positive LR was 25.9 (3.7%-180.7%), and a low negative LR was (0.03 [CI, 0.00% to 0.17%]), indicating high accuracy for urethral specimens.


Diagnostic performance of Gram-stained endocervical, urethral swabs and urine smears

Among the six datapoints, the pooled sensitivity was 81% (CI, 67%-90%) and the pooled specificity was 98% (CI, 93%-99%). The positive LR (40.8 [CI, 11.5%-143.8%]), negative LR (0.19 [CI, 0.10%-0.36%]), and DOR (215 [CI, 41%-1126%]), indicated reduced accuracy for specimens other than urethral specimens.


The present meta-analysis suggests that GSS has the highest accuracy when compared to reference culture methods; however, its accuracy is relatively less than that of NAAT (Gen-Probe, Ng-PCR Rotor gene system, and Xpert® CT/NG assay). Additionally, GSS showed the highest accuracy in the urethral compared to other specimens, such as endocervical, urethral swabs, and urine specimens. However, all subgroups showed high positive LRs, low negative LRs, and high DORs, and the best LRs and DORs were reported for GSS compared to culture methods, followed by those of GSS compared to NAAT. In contrast, the best LRs and DORs were reported for Gram-stained urethral smears, followed by those of Gram-stained endocervical, urethral swabs, and urine smears. Given the convenience of GSS and their rapid turnaround time, these results show great potential for expanded first-line screening for Ng infection and demonstrate the utility of GSS in gonorrhea screening of at-risk populations.


The high positive and low negative LRs found in each subgroup, especially those that tested urethral specimens and those obtained after comparing GSS to reference methods, also imply that GSS can meaningfully inform the posttest probability of infection. The pooled accuracies of these subgroups have implications for their use in clinical and nonclinical outreach settings. For example, Gram-stained endocervical, urethral swabs, and urine smearsshowed a slightly higher false-negative rate than Gram-stained urethral smears.[22,23] The false-negative rate is of particular concern in high-risk groups, in which a high rate is more likely to lead to an undetected infection. In such scenarios, timely confirmatory testing could resolve the preliminary screening results. However, the convenience and rapid turnaround time of GSS and their ease of use compensate for their slightly lower sensitivity. In summary, the GSS can be safely integrated into expanded screening initiatives as a first-line screening test.


The results of this study should be interpreted with caution. First, reference standards were found to influence the accuracy of GSS.[22,24] When the ideal culture reference standard was used, specificity was higher than when a NAAT reference standard was used.[25] In contrast, thesensitivity was high when a NAAT reference standard was used.[26] Only seven of the included studies[11,14,16,17,18,19,20] used the culture reference standard to ascertain true disease status. Misclassification by reference standards is known to influence the measured sensitivity and specificity of the index tests.[23] Accuracy estimates from studies that used imperfect reference standards to ascertain true disease status may have been artificially inflated (as more sensitive than culture reference standard) or lowered (does not allow for testing of antimicrobial susceptibility) because of misclassification by the reference standard. Standardization of reference standards is required for future diagnostic accuracy studies.


Second, the effect of antimicrobial susceptibility testing on diagnostic accuracy is worth further consideration. Antimicrobial susceptibility testing leads to the expansion of surveillance of antimicrobial resistance and treatment failures, and promotes responsible antimicrobial use and stewardship.[27]


Third, the index tests included in this meta-analysis detected intercellular diplococcus Ng and therefore could not detect infection within 2-10 days.[28] If the clinical suspicion of a positive GSS result is high, further testing is required. In the case of a possible false-negative result, further screening with another conventional laboratory-based test could be considered, depending on available resources. More research is needed to determine how to effectively link screening with further linkages and follow-up, especially in hard-to-reach populations and low-resource settings.


Finally, evidence on GSS will be of greater use to policymakers and guideline developers if outcomes are documented beyond accuracy. These include patient-centered outcomes and operational research outcomes, such as acceptability, preference, feasibility, and impact. Future research on the cost-effectiveness of GSS in different settings, populations, and contexts is warranted to make informed decisions on this test and on testing strategies.


GSS offers many advantages: a fast turnaround time,[29,30] declaration of results at the point of care with the potential for affecting clinical management, early detection of undiagnosed cases of gonorrhea,[29,30] and high intra and inter-observer agreement or concordance, and reduction in the test cost.[31] Given the lack of global evidence, this review comes closer to independently assessing the role of GSS for widespread use in the field by synthesizing all available data on their accuracy and providing further evidence on the benefits of GSS.


In this study, GSS is found to be accurate and suitable for screening initiatives. As a result of their accuracy and the urgent need to increase gonorrhea screening in marginalized and at-risk populations, this test could play a substantial role in expanded screening initiatives, which would eventually impact the control of gonorrhea infection at the population level.


Funding

None.


Ethics approval and consent to participate

Not applicable.


Conflicts of interest

The authors declare no conflict of interest.


  1. Młynarczyk-Bonikowska B, Majewska A, Malejczyk M, Młynarczyk G, Majewski S. Multiresistant Neisseria gonorrhoeae: a new threat in second decade of the XXI century. Med Microbiol Immunol 2020;209:95-108 DOI: 10.1007/s00430-019-00651-4
  2. Rubin DHF, Ross JDC, Grad YH. The frontiers of addressing antibiotic resistance in Neisseria gonorrhoeae. Translational Research: the Journal of Laboratory and Clinical Medicine 2020;220:122-37 DOI: 10.1016/j.trsl.2020.02.002
  3. Unemo M, Seifert HS, Hook EW, Hawkes S, Ndowa F, Dillon J-AR. Gonorrhoea. Nat Rev Dis Prim [Internet] 2019;5:79. DOI: 10.1038/s41572-019-0128-6
  4. Meyer T, Buder S. The Laboratory Diagnosis of Neisseria gonorrhoeae: Current Testing and Future Demands. Pathogens 2020;9:91. DOI: 10.3390/pathogens9020091
  5. Qian S, Foster R, Bourne C, Vickers T, McIver R, McNulty A. Neisseria gonorrhoeae positivity in clients presenting as asymptomatic contacts of gonorrhoea at a sexual health centre. Sex Health 2020;17:187-191. DOI: 10.1071/SH19091
  6. Thairu Y, Nasir I, Usman Y. Laboratory perspective of gram staining and its significance in investigations of infectious diseases. Sub-Saharan African J Med 2014;1:168-174. DOI: 10.4103/2384-5147.144725
  7. Papp JR, Schachter J, Gaydos CA, Van Der Pol B. Recommendations for the laboratory-based detection of Chlamydia trachomatis and Neisseria gonorrhoeae—2014. MMWR Recomm reports Morb Mortal Wkly report Recomm reports/Centers Dis Control. 2014;63:1.
  8. Whiting PF, Rutjes AWS, Westwood ME, Mallett S, Deeks JJ, Reitsma JB, et al. QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies. Ann Intern Med 2011;155:529-536. DOI: 10.7326/0003-4819-155-8-201110180-00009
  9. Harbord RM, Whiting P. Meta-analysis of diagnostic accuracy. In: Sterne JA, ed. Meta-analysis in Stata: An Updated Collection From the Stata Journal. College Station, TX: Stata Pr; 2009.
  10. Jaeschke R, Guyatt GH, Sackett DL. Users’ guides to the medical literature. III. How to use an article about a diagnostic test. B. What are the results and will they help me in caring for my patients? The Evidence-Based Medicine Working Group. JAMA 1994;271:703-707. DOI: 10.1001/jama.271.9.703
  11. Bhargava A, Bala M, Singh V, Joshi NC, Kakran M, Puri P, et al. How reliable is microscopy and culture for the diagnosis of gonorrhea? An 11-year experience from India. Sex Transm Dis 2017;44:111-113. DOI: 10.1097/OLQ.0000000000000550
  12. Situ SF, Ding CH, Nawi S, Johar A, Ramli R. Conventional versus molecular detection of Chlamydia trachomatis and Neisseria gonorrhoeae among males in a sexually transmitted infections clinic. Malays J Pathol 2017;39:25-31.
  13. Hananta IPY, Van Dam AP, Bruisten SM, Van Der Loeff MFS, Soebono H, Christiaan De Vries HJ. Value of light microscopy to diagnose urogenital gonorrhoea: A diagnostic test study in Indonesian clinic-based and outreach sexually transmitted infections services. BMJ Open 2017;7:1-12. DOI: 10.1136/bmjopen-2017-016202
  14. Goodhart ME, Ogden J, Zaidi AA, Kraus SJ. Factors affecting the performance of smear and culture tests for the detection of Neisseria gonorrhoeae. Sex Transm Dis 1982;9:63-69. DOI: 10.1097/00007435-198204000-00002
  15. Juchau S V., Nackman R, Ruppart D. Comparison of gram stain with DNA probe for detection of Neisseria gonorrhoeae in urethras of symptomatic males. J Clin Microbiol 1995;33:3068-3069. DOI: 10.1128/jcm.33.11.3068-3069.1995
  16. Taylor SN, Dicarlo RP, Martin DH. Comparison of methylene blue/ gentian violet stain to gram’s stain for the rapid diagnosis of gonococcal urethritis in men. Sex Transm Dis 2011;38:995-996. DOI: 10.1097/OLQ.0b013e318225f7c2
  17. Goh BT, Varia KB, Ayliffe PF, Lim FK. Diagnosis of gonorrhea by gram-stained smears and cultures in men and women: role of the urethral smear. Sex Transm Dis 1985;12:135-139. DOI: 10.1097/00007435-198507000-00008
  18. D’Angelo LJ, Mohla C, Sneed J, Woodward K. Diagnosing gonorrhea. A comparison of standard and rapid techniques. J Adolesc Heal Care 1987;8:344-348. DOI: 10.1016/0197-0070(87)90006-4
  19. Orellana Miguel MA, Gómez-Lus ML, Lora D. Sensitivity of Gram stain in the diagnosis of urethritis in men. Sex Transm Infect 2012;88:284-287. DOI: 10.1136/sextrans-2011-050150
  20. Bartelsman M, Straetemans M, Vaughan K, Alba S, Van Rooijen MS, Faber WR, et al. Comparison of two Gram stain point-of-care systems for urogenital gonorrhoea among high-risk patients: Diagnostic accuracy and cost-effectiveness before and after changing the screening algorithm at an STI clinic in Amsterdam. Sex Transm Infect 2014;90:358-362. DOI: 10.1136/sextrans-2013-051500
  21. Borg S, Daniel J, Forsyth S. Is urethral microscopy useful in symptomatic male patients without urethral discharge? Int J STD AIDS 2019;30:79-81. DOI: 10.1177/0956462418793511
  22. Ng LK, Martin IE. The laboratory diagnosis of Neisseria gonorrhoeae. Can J Infect Dis 2005;16:15-25. DOI: 10.1155/2005/323082
  23. Whiting P, Rutjes AWS, Reitsma JB, Glas AS, Bossuyt PMM, Kleijnen J. Sources of variation and bias in studies of diagnostic accuracy: a systematic review. Ann Intern Med 2004;140:189-202. DOI: 10.7326/0003-4819-140-3-200402030-00010
  24. Verma R, Sood S. Gonorrhoea diagnostics: An update. Indian J Med Microbiol 2016;34:139-145. DOI: 10.4103/0255-0857.180278
  25. Budkaew J, Chumworathayi B, Pientong C, Ekalaksananan T. Conventional culture versus nucleic acid amplification tests for screening of urethral Neisseria gonorrhea infection among asymptomatic men who have sex with men. Pragmatic Obs Res 2017;8:167-73. DOI: 10.2147/POR.S137377
  26. Ghanem KG, Hynes NA, Deputy D, Bloom A. Clinical manifestations and diagnosis of Neisseria gonorrhoeae infection in adults and adolescents. UpToDate; 2019.
  27. Unemo M, Seifert HS, Hook EW, Hawkes S, Ndowa F, Dillon J-AR. Gonorrhoea. Nat Rev Dis Prim 2019;5:79. DOI: 10.1038/s41572-019-0128-6
  28. Whittles LK, White PJ, Didelot X. A dynamic power-law sexual network model of gonorrhoea outbreaks. PLoS Comput Biol 2019;15:e1006748-e1006748. DOI: 10.1371/journal.pcbi.1006748
  29. Jordan SJ, Schwebke JR, Aaron KJ, Van Der Pol B, Hook 3rd EW. Meatal Swabs Contain Less Cellular Material and Are Associated with a Decrease in Gram Stain Smear Quality Compared to Urethral Swabs in Men. J Clin Microbiol 2017;55:2249-2254. DOI: 10.1128/JCM.00423-17
  30. Raffe S, Soni S. Diagnostic tests for sexually transmitted infections. Medicine Baltimore 2018;46:277-282. DOI: 10.1016/j.mpmed.2018.02.008
  31. Sarier M, Sepin N, Duman I, Demir M, Hizel A, Göktaş Ş, et al. Microscopy of Gram-stained urethral smear in the diagnosis of urethritis: Which threshold value should be selected? Andrologia 2018;50:1-4. DOI: 10.1111/and.13143